Figure 6-48 Wiring Diagram: Incubator Relay Board 9 Or Lower - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

Chapter 6: Illustrated Parts
To J33 on
Relay Board
To J30 on
Relay Board
Note:
The two position connector connects to the two position connector on
harness 6600-1535-700. See Radiant Warmer section of Wiring Diagram.
The configuration plug is always 6600-0748-701 (115 VAC).
See Note
1
Above
2
To J39 on
red
black
Relay Board
Humidifier Add Water Thermostat and Bottle
Position Sensor
To J32 on
Relay Board
6600-0730-700
Ground Equalization Post
226
red
GND
black
GND
brown
orange
Air flow signal
shield
+5V
6600-0729-700
6600-0711-700
black
brown
red
orange
yellow
green
blue
violet
brown
red
6600-1003-600
Solid State Relay
+12V
+
AC
Htr sig
U2
-
AC
6600-0864-700
red
RHADDWATER
brown
RHBOTTLE
black
GND
Ground Wires
6600-0870-700

Figure 6-48 Wiring Diagram: Incubator Relay Board 9 or Lower

6600-0343-000 104
Incubator
5
5
4
4
3
3
2
2
1
1
Optical Fan Sensor
6600-0820-700
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
See Note
yellow
Above
blue
orange
white
blue
black
6600-0803-700
red
1
1
brown
2
2
black
3
3
shield
shield
4
4
6600-0863-700 (X2)
Bed Chassis
Cover Ground
6600-0748-702
6600-0748-701
230V
Configuration Plug Group 1
6600-1057-600
A
M
B
Inc.
C
Brushless
Motor
6600-1041-700 x2
2
white
1
1
white
2
2
2
white
1
1
white
2
2
Heater (X2)
Tstat
6600-1019-600
6600-0709-700
Add Water Tstat
1
2
Rh Bottle Open Sw
Heatsink Ground
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
115V
1
1

Advertisement

Table of Contents
loading

Table of Contents