Chapter 1: Functional Description
The RS-485 integrated circuit converts the RS-232 TTL signals from the microcontroller to RS-485 signals for the
bus.
The timekeeping RAM has a battery integrated into the chip so that the time and date run are kept current
even with the power off. The battery has a minimum life of 10 years. Operator configuration values, patient
information (trending), and the error log are stored in this RAM.
The graphics controller is an S-MOS VGA controller. The graphics controller interfaces the data from the video
RAM to the EL display. The controller also synchronizes the display using a horizontal pulse (LP) and a vertical
pulse for the whole display frame (YD). The controller handshakes with the 80C188 using the READY line to
eliminate any lost data during display refreshes.

1.4 LED Board

The LED Board contains five display banks and two display drivers. One of the display drivers controls the
patient temperature and air temperature display banks. The other driver controls the patient set temperature,
air set temperature, warmer bar graph, and the mode and override indicators.
This allows the two large displays (patient and air temperature) to be multiplexed at a slower rate than the
other LEDs. This results in brighter large displays. Each driver has a brightness potentiometer that is preset at
the factory and should not be adjusted in the field.

1.5 Power Supply

The universal input switching power supply converts the line voltage to +5V DC and +12V DC. This supply can
source up to 75 watts. The 5 volts powers the electronics and the 12 volts is used by the EL display and for
future boards. Acceptable voltage ranges are 4.75V - 5.25V and 10.8V - 13.2V.

1.6 Peripheral Components

For the heater isolation transformer with relay board revision 9 or lower, the heater isolation transformer
isolates the incubator and radiant heaters from the mains voltage. The primaries mains input voltage comes
from the relay board and is switched by safety relay 1 and is also routed through the heater current sense
circuit. The dual primaries are connected in parallel for 115 volt operation and in series for 230 volt operation.
The isolated secondary output voltage is always 115 volts and goes directly to the solid state relays and the
incubator and radiant heaters.
For the heater isolation transformer with relay board revision 10 or higher, the heater isolation transformer
isolates the incubator and radiant heaters from the mains voltage. The primaries mains input voltage comes
from the relay board and is switched by safety relay 1. The dual primaries are connected in parallel for 115 volt
operation and in series for 230 volt operation. The isolated secondary output voltage is always 115 volts and
is input to the relay board where it is used by the heater current sense circuit. It then outputs from the relay
board to the solid state relays and the incubator and radiant heaters
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
5
6600-0343-000 104