Chapter 2: Service Checkout
7. Check the canopy seals:
a. If the unit has original canopy seals, which are attached with fasteners, check for visible cleanliness.
b. If the unit has enhanced canopy seals, which can be removed by hand, check the following:
Check for visible cleanliness.
Make sure that all seals (the two south corner covers, the two long side seals, the short south seal,
and the notched north seal) can be easily removed and reinstalled. For details, refer to section 5.2.1,
steps 10 and 15.
Make sure that none of the red indicators are visible after reinstallation.
8. Check the portholes. Open the portholes by pressing on the latch. The cover should swing open. Close the
porthole and check that the latch holds the cover securely shut and that the cover seals tightly against the
porthole gasket. Check that all the porthole seals are in place and are in good condition.
9. Check that the tubing access covers in the four bed corners and the large slot grommet at the head of the
bed are in place and are in good condition.
10. Check the operation of the bed. The bed should rotate easily without binding. If the bed is properly seated
and locked in place, the mattress should be level. With the bed rotated back into the straight position,
check to see that the bed platform extends and stops when it is pulled out on either side. Check the
operation of the bed tilt mechanism. Squeeze the tilt control and push down on the foot of the bed. The
head of the bed should raise easily, and should stay in position at any angle along its tilt path when the tilt
control is released. Push down on the head of the bed. The foot of the bed should raise easily, and should
stay in position at any angle along its tilt path when the tilt control is released.

2.3 Controller Checks

WARNING:
Do not use the OmniBed in the presence of flammable anesthetics: an explosion hazard exists
under these conditions.
1. Connect the OmniBed power cord to a properly rated outlet.
2. Connect the patient probe to jack 1 on the probe panel.
3. Switch on the power at the mains switch on the back of the unit, and at the standby switch on the jack
panel, while holding in the override button (>37) on the control panel during power up until the software
revision screen appears. Release the button and the first service screen will appear.
4. Scroll to "Down" and select it to bring up the second service screen. Select Status to see Status screen.
Check status of the software self tests. These include: incubator heater on (INCHTRON), warmer heater on
(WRMHTRON), incubator/warmer heater off (I/WHTROFF), humidifier heater on (RHHTR), remote monitoring
data stream (RS232LOOP), incubator fan on (FANON), and incubator fan off (FANOFF). All test should say
PASS except RS232 LOOP (the connector pins must be shorted to get the PASS message).
12
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.