Chapter 4: Troubleshooting
Cause(s)
Defective relay board.
Defective control board.

4.4.27 System Failure 33

System failure 33 indicates: Defective RH Solid State Relay.
Cause(s)
Defective RH Solid State Relay.

4.4.28 System Failure 34

System failure 34 indicates: Software revision level does not match.
Cause(s)
The software revision of the control board and the
display board are not compatible

4.4.29 System Failure 35

System failure 35 indicates: Wrong dipswitch set up.
Cause(s)
Switches set wrong
Defective control board. Replace control board.

4.4.30 System Failure 36

System failure 36 indicates: LV1 differs from LV2 by >10% of LV2.
Cause(s)
Mains Voltage Comp Circuit is not calibrated correctly.
Defective 50 pin ribbon cable between the relay and
control board.
52
Action(s)
Check the 8 position dipswitch on the control board. Verify that Switch 1 and Switch 2 are on.
6600-0343-000 104
Action(s)
Measure the voltages at the relay board:
J37 pin 48-45 and J37 pin 49-45.
If mains voltage 100/115 = 3-5 volts, difference
within 0.4 volts.
If mains voltage 230 = 7-10 volts, difference within
0.8.
If not, replace the relay board.
If the voltages are OK replace the control board.
Action(s)
Replace relay board
Action(s)
This error may occur after replacing the control/display board.
Be sure to install the software provided in the control/display
board. Replace the System Software.
Action(s)
Calibrate the line voltage.
Check for continuity on pin 48 and 49 (LineComp1,2) If
defective replace cable.