Caster Replacement; Humidifier Assembly Repairs - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

5.11 Caster Replacement

Refer to "Figure 5-53 Elevating Base".
The casters may be replaced with the unit upright or carefully placed on its back. You may wish to lock the
canopy in its shipping position.
1. Lock all the other casters.
2. Lift the caster off the floor and use blocks to support the leg near the caster you are replacing.
3. On newer casters with mounting plates, remove screw from boss underneath the leg casting.
On older casters with a set screw on the side of the leg casting, remove the screw on the side of the leg
that holds the caster in its mounting hole. Remove the caster.
NOTE: The new style caster with a mounting plate has a different shaft diameter than the older caster and
will not fit on the older set screw style leg casting.
4. On newer casters, secure the caster to the leg by tightening the screw through the caster plate into the
boss in the bottom of the leg.
5. On older casters, line up the hole in the caster stem with the set screw hole in the leg and insert the caster.
Apply Loctite #242 to the screw. As you tighten the screw the caster should be drawn into the leg.
NOTE: If you are replacing a caster hub cover, you must use an 1/8" drill to remove the old cover stem before
snapping the new hub cover in place.

5.12 Humidifier Assembly Repairs

Refer to "Figure 5-54 Chassis Bottom Cover" and "Figure 5-56 Humidifier Parts".
A design change was made to the humidifier assembly to make it easier to remove the screws during
disassembly. Units shipped since about March 2002 have this revised design. The earlier design can be
identified by the button head cap screws used in the assembly. The later design uses Phillips head screws to
make disassembly easier.
When working on the earlier design if you have difficulty with disassembly you may wish to replace the
entire humidifier assembly with the later design 6600-0245-850 . When disassembling the older design, we
recommend that you replace the old screws with the new screws listed below. You may also require the other
parts listed below if they are damaged during disassembly. (Refer to section 5.12.1.)
The reservoir switch button was also redesigned to prevent it jamming in the closed position. You should also
replace this button when working on an old style humidifier.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures
6600-0343-000 104
133