Heater Door Cable Adjustment - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

3. To replace the door cable, use a flat screwdriver blade to pry the top tang (12 o'clock position) on the spool
assembly and pull the cable off the spool toward you. Use a 2.5 mm hex key to separate the cable from the
spool.

5.6.4 Heater Door Cable Adjustment

Refer to "Figure 5-44 Heater Door Cable Adjustment".
1. The bottom end cap on the right side upright should be removed. Position the top nut and flat washer so
the bottom of the flat washer is in the middle of the threads. Slide the heater door opening rod through the
bracket, then install the flat washer, lock washer, and nut (8mm wrench).
2. Use a straight blade screwdriver in the slot at the cable rod's end to hold the rod and use the wrench to
tighten the bottom nut.
3. Turn on the unit and run the canopy to its upper travel limit, and verify that both doors fully open.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures
6600-0343-000 104
117

Advertisement