Figure 4-2 Second Service Screen; Figure 4-3 Status Menu - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

Scale R
Select from 2 scale resolution settings; 10 grams or 5 grams (10 grams is the factory default).
Comfort
Allows you to remove the Comfort Screen feature.
Set real time clock for time, day and date. Choose how date is displayed (North American or European).
Set Time
Choose a 12 hour (AM/PM) or 24 hour time display.
Display the current software revision of the options installed on this specific unit (Humidifier, Scale, SPO2,
View Mods
etc.)
Down
Go to second service screen.
Up
Return to previous screen.
Check status of all self test the software
runs continuously. These include: incubator
heater on (INCHTRON), warmer heater on
(WRMHTRON), incubator/warmer heater
off (I/WHTROFF), humidifier heater on
(RHHTR), remote monitoring data stream
(RS232LOOP), incubator fan on (FANON), and
Status
incubator fan off (FANOFF).
If the RS232 option is not installed
RS232LOOP will display N/A. To test the
circuit if the option is installed, short pins 2
and 3 on the 9 pin connector on the back of
the electrical enclosure.
Select to bring up a diagnostic diagram of the unit that displays the status of all the switches. If the circle
next to the switch is lit, the switch is closed; if it is not lit, the switch is open. Also, while the switch status
Switches
screen is active, you can hold down the alarm silence button to light the alarm light, system failure light
and all LED segments to test them.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting

Figure 4-2 Second Service Screen

Figure 4-3 Status Menu

6600-0343-000 104