Figure 6-45 Wiring Diagram: Electrical Enclosure - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

6600-0736-702
Left/West
Hood Hand Control
6600-0736-701
Right/East
Hood Hand Control
6600-0746-700
Ebase / Hood
Transformer
Standby Switch
6600-0227-850
6600-0800-700
Outlets
6600-0583-600 (X3)
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Hood Hand Control
UP
1
1
2
2
DOWN
3
3
4
4
5
5
UP
1
1
2
2
DOWN
3
3
4
4
5
5
Ebase / Hood Transformer
Configuration Plug Group 1
Standby Switch
BATTERY
1
3
BT1
6600-1024-600
2
4
Power Supply
Power Supply
6600-0708-700
6600-0221-850
+5 V
+5V
PHASE
+5V
GND
NEUTRAL
GND
GND
AC GND
+12V
+12 V
Mains Outlets
P
G
N
1
P
G
1
N
6600-0562-603 X2
P
Circuit Breakers 3.5A.
G
N
To J51 on
Relay Board

Figure 6-45 Wiring Diagram: Electrical Enclosure

6600-0744-700
UPHOOD
DOWNHOOD
GND
UPHOOD
DOWNHOOD
GND
Sec A 14.6vac
Sec A 14.6vac
Sec B 14.6vac
Sec B 14.6vac
6600-0748-702
6600-0748-701
230V
Configuration Plug Group 1
V charge out
V charge in
V battery
GND
6600-0707-700
red
white
blue
green
black
brown
orange
yellow
red
black
2
3
1
4
2
2
Power Switch/
Circuit Breaker
6600-1014-601 (9 Amp, 230V)
6600-1014-602 (12 Amp, 115V)
brown
6600-0731-700
blue
includes wiring to outlets
6600-0343-000 104
Chapter 6: Illustrated Parts
red
brown
To J35 on
black
black
Relay Board
red
brown
red
To J44 on
white
violet
Relay Board
black
To J52 on
Phase
black
Neutral
white
Relay Board
115V
To J43 on
red
black
Relay Board
red
black
To J47 on
Phase in
red
Phase out
black
Relay Board
When checking for
the secondary voltage
5 and 12 VDC from
the power supply to
the relay board J 41,
make sure the power
To J41 on
cord is connected
and both the breaker
Relay Board
on the back and the
switch on the front
To J48 on
panel are in the ON
Relay Board
position.
P
Line Filter
G
N
6600-1006-600
120VAC into relay board and out
to power supply. If in good and
out missing check fuses on relay
board.