Tips For Repairing An Old Style Unit; Disassembly - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

Chapter 5: Repair Procedures
Qty. per assy.
3
4
4
2
4
1
1
1

5.12.1 Tips for Repairing an Old Style Unit

Some of the hex head screws may be difficult to remove. To minimize the chances of stripping the heads, be
sure that the 1.5 mm Allen key you are using is not worn, and is not a ball head style. If you strip the heads,
you may need to drill them out. If you drill out the heads, you may need to replace the ramp block and/or the
heater mount if you have difficulty in removing the body of the screw.

5.12.2 Disassembly

1. Remove the humidifier reservoir.
2. Remove the 6 screws that secure the wire raceway cover and remove the cover.
3. Remove the chassis cover. (Refer to section 5.9.1.)
4. Disconnect the 3 electrical connectors.
5. Remove the 2 screws on either side of the heater assembly.
6. Remove the heater assembly.
7. Use a 2mm hex key to remove the 2 screws from the reservoir switch and remove the switch.
8. Remove the 4 screws that secure the top bracket. Back the bracket off feeding the wire harnesses through
as you remove it.
9. Remove the top gasket, feeding the wire harnesses through as you remove it. The thermostat wire harness
can slide through the slots in the gasket.
10. Remove the 4 screws in the heater mount and remove the bottom bracket.
11. Remove the socket head cap screw next to the add water thermostat, then remove the bottom gasket,
insulating cylinder, and protective insert.
134
Description
Screw, M3 X 8, Phillips head, Teflon coated
Screw, M3 X 10, Phillips head, Teflon coated
Screw, M3 X 12, Phillips head
Screw, M3 X 16, Phillips head
Spacer, .125 ID
Ramp block
Heater mount
Button, reservoir switch
6600-0343-000 104
Part Number
6600-1255-401
6600-1255-402
6600-1255-403
6600-1255-404
6600-1779-500
6600-1777-500
6600-1291-500
6600-1298-500
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.