6 Postprocessing

6.1 After each treatment session

14
6
Postprocessing
6.1
After each treatment session
CAUTION!  Risk of contamination! Perform reprocessing immediately
after treatment, after 1 hour at the latest.
ü The preparation instrument is at a standstill.
ü Wear suitable protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the preparation instrument with tweezers.
3. Detach the instrument from the motor.
4. Transport the instrument to the hygiene room in a suitable transport
container.
5.
CAUTION!  Risk of contamination! Conduct hygienic
reprocessing  [→ 15]. Observe the relevant national/local
regulations.
6.2

At the end of the work day

NOTICE!  Do not leave any instruments on the motor overnight, in order
to prevent oil from leaking into the electric motor. Never lubricate the
electric motor.
Dentsply Sirona
®
Instructions for use Midwest
T1/T2 Energo™
68 19 200 D3751
D3751.201.03.02.02    2023-11