Switching The Unit On And Starting The Software; Calibrating The Motor Arm - Dentsply Sirona inEos X5 Operating Instructions Manual

Hide thumbs Also See for inEos X5:
Table of Contents

Advertisement

6 Initial startup

6.4 Calibrating the motor arm

28
6.3.4

Switching the unit on and starting the software

NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
ü Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry.
Ä The unit is dry and can be put into operation.
1. Switch the unit on at the main switch.
2. Start the software at the PC via the Windows start menu.
6.4
Calibrating the motor arm
IMPORTANT
Calibrating the system
Following installation of the inEos X5, it is recommended that you
perform an axis calibration with the inLab SW software.
ü The inEos X5 is switched on.
ü The jaw model supplied for calibration is ready.
ü You have restarted the software.
1. Click the "Devices" button in the system menu.
2. Click on "inEos X5" .
Ä A selection menu opens.
3. Click on the "Calibrate MotorArm" button.
Ä The calibration dialog opens.
4. Follow the steps in the calibration dialog.
5. Position the calibration model on the rotation disk in such a way that
the front teeth are facing directly in the direction of the unit.
6. Start the scan process.
Ä The unit is calibrated. Wait until the process has been
completed.
7. Restart the software.
Dentsply Sirona
Operating Instructions inEos X5
64 17 096 D3586
D3586.201.01.16.02    2023-12

Advertisement

Table of Contents
loading

Table of Contents