Disinfection By Immersion (Chemical) - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

6.2.3 Disinfection by immersion (chemical)

If a steam disinfection chamber is not available, the dismantled parts can be chemically
disinfected by means of immersion.
Immerse the dismantled components in the solution with the disinfectant, following the
manufacturer's instructions.
When disinfection is complete, rinse with running, preferably decalcified, water; shake and
drain off any remaining water. Leave the components to dry completely.
After this first operation and before placing the components in an autoclave, wrap them in
muslin or in a similar material.
Effective sterilization is achieved in an autoclave at 121°C for approx. 15 minutes.
6-4
Do not disinfect, sterilize or re-use disposable products.
Disinfect and sterilize every time an infected patient is ventilated.
In normal conditions, disinfect and sterilize depending on the frequency
of the lung ventilator's use and in any case at least once a month.
That the components should be checked every time they are sterilized
and any damaged parts should be replaced.
Carrying out a functioning test of the machine whenever parts or
components are replaced.
Not using formaldehyde or phenol-based disinfectants as they can
cause cracking and reticulation of plastic parts;
Not using too strong disinfectants as they can compromise the working
life of the immersed parts;
Rinsing and carefully drying the components since marks and other
damage can occur when the components are exposed to high
temperatures.
Always refer to the instructions provided by the autoclave manufacturer
regarding temperature and time.