3.5.3 Alarm limits check

The alarms must trigger at the proper time and in the correct manner.
Before starting the alarm limits, check the alarms setup, and change the set
values whenever necessary (please see previous paragraph).
High Pressure
PAW
Low Pressure
3-32
WARNING !! Severe patient injuries
Check the proper activation of the visual and acoustic signals.
Set the high pressure alarm limit to a value higher than the PAW
ventilation pressure by 5 cmH
Block the patient simulator (using your hands) during ventilation.
The system activates the airways HIGH PRESSURE alarm: silence
the alarm.
Unlock the patient simulator.
Set the low pressure alarm limit to 5 cmH
Disconnect the patient circuit during ventilation.
After about 20 seconds the system activates the airways LOW
PRESSURE alarm: silence the alarm.
Reconnect the patient simulator.
10. Select VC/VAC operative
mode.
11. Press START: the lung
ventilator begins its ventilation
cycle.
O.
2
O.
2