Man Operative Mode; Apnoea Back-Up - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

4.6.12 MAN operative mode

MANUAL ventilation available in all operative modes.
By selecting the MAN mode (at the bottom of the display) the system provides the patient
with a breath.
The breath ventilation parameters depend on the set operative mode. The function
activation is monitored by the screen and signalled by the green LED inside the box, that
turns on. This mode is active while the lung ventilator is running.

4.6.13 APNOEA BACK-UP

Apnoea BACK-UP is a safety function available in two of the Operative modes: PSV,
PSV-TV and CPAP. Apnoea BACK-UP function enables if the patient, ventilated in one
of the modes above, stops breathing.
After a pre-set time (Apnoea Time) in the SETUP - Other, will appear the relative alarms
and the system automatically starts ventilating the patient.
4-42
In APRV the patient breaths on two positive pressure levels not
synchronised with the patient's spontaneous breathing.
WARNING !! Patient injury hazard
If the patient stops breathing, after 20 seconds the system enables the low
PAW acoustic and visual alarm (low airway low pressure).
WARNING !! Patient injury hazard
When the lung ventilator switches to this safety mode automatically, the
User CANNOT edit the ventilation parameters.
The lung ventilator continues its activity and the User acknowledges the
emergency condition.
When the apnoea BACK-UP function enables, the ventilation
parameters used are those set based on the selected operative mode.
User manual, DU3104101

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents