3.2.8 Data Connection (Trend and Events downloading)

3-12
Verify the functioning of the Mainstream GAS ANALYZER
Switch ON the electric power supply.
A green LED (on the gas analyzer) indicates that the IRMA analyzer is
ready for use.
Verify the functioning of the Sidestream GAS ANALYZER
Switch ON the electric power supply.
A green LED (on the gas analyzer - sampling line) indicates that the ISA
analyzer is ready for use.
GAS ANALYZER: Zeroing
Mainstream IRMA sensor: after around one minute from the device is
turned on, it is necessary to perform the manual procedure of Zero
Calibration (please see on chapter 4: SETUP MENU).
With sidestream ISA sensor: the zeroing of gas sensor measurement is
activated in automatic mode; anyway, it is possible to perform the Zeroing
calibration procedure when needed (the green led blinks when the GAS
ANALYZER turn-on).
Connect a USB flash drive to the USB1 socket used for CPU
programming and data download (USB socket near the reset button).
For more information see chapter 4.2.2 (SETUP - Other: Save to USB).
Request the guide with the relative instructions for data interpretation
(Procedure for downloading data).