3.2.3 Medical gas connection

Screw the pneumatic O
lung ventilator connector.
Connect the pneumatic O
the relevant connector on the gas system.
Be sure that the pneumatic hose is well fixed
and that the medical gas system is clean and
without lubricants.
3-8
The fast connections mounting, suitable for the medical gas distribution
system, as well as all pneumatic supply hoses maintenance and/or
replacement operations should be carried out only by qualified staff, so as
to prevent any gas inversions that might be fatal for the patient.
supply hose on the
2
supply hose on
2
PNEUMATIC SUPPLY
The lung ventilator is supplied with a pneumatic hose including a DISS
(Diameter Index Safety System). The installation technician taking care of
the hose connection to the outlet must ensure that they are compatible with
the hospital medical gas pipeline system.
WARNING !! Unit failure risk
In order that the lung ventilator operates as specified, the input medical
gas pressure should be between 280 kPa and 600 kPa (2.8 - 6 bar / 40
- 86 psi).
Before using the lung ventilator, make sure that this requirement is met.
After connecting the medical gas supply hose, make sure that the
system works properly.
The lung ventilator can work even without an oxygen supply using the
internal turbine only; in this case the FiO
WARNING !! Patient injury hazard
The oxygen should be of medical type, without oil and filtered.
The medical gas source shouldn't contain water: if you suspect the
presence of water, connect a water trap to avoid damages on the lung
ventilator and its components.
could be set only to 21%.
2
User manual, DU3104101