A.6.4 Table 4

Recommended separation distances between portable and mobile RF communications
equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the device as recommended below, according
to the maximum output power of the communications equipment.
Rated
Separation distance according to frequency of transmitter (m)
maximum
output power
From 150kHz to 80MHz
of transmitter
( W )
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Notes:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
A-28
From 80MHz to 800MHz
d = 1,2 ⋅√P
0,12
0,38
1,2
3,8
12
From 800MHz to 2GHz
d = 1,2 ⋅√P
0,12
0,38
1,2
3,8
12
d = 2,3 ⋅√P
0,23
0,73
2,3
7,3
23