1-Point Fluoride Sensor Calibration (Cal) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller dialog daca
Table of Contents

Advertisement

Calibration
CAL F -
Two point calibration
Immerse sensor in buffer 2
Sensor value
Sensor voltage
Start with <CAL>
Fig. 53: Fluoride sensor calibration (CAL)
8.
Immerse the sensor in test con‐
tainer 2 with calibration solution.
When doing so gently move the
sensor
9.
Then press
[Calib. in progress]
ð
10.
Then press
value or press
the calibration
11.
Then press
12.
Import the result of the calibration
into the controller memory by
pressing the
ð The controller displays the con‐
tinuous display again and oper‐
ates with the results of the cali‐
bration.
82
4.88 ppm
144.2 mV
A1041
.
to adjust the ppm
to continue with
key
Should the result of the cal‐
ibration lie outside the
specified tolerance limits,
an error message appears.
In this case the current cal‐
ibration will not be applied.
Check the prerequisites for
the calibration and clear
the error. Then repeat the
calibration
1-point fluoride sensor cali‐
9.3.3
bration (CAL)
Correct sensor operation
Correct measuring and metering
is only possible if the sensor is
working perfectly
Observe the sensor operating
instructions
The carrying out of a 2-point cali‐
bration is strongly recommended
and is to be preferred to other
methods
For calibration the sensor must
be removed and refitted in the in-
line probe housing. To do this,
observe the operating instructions
of your in-line probe housing
Material required for calibration of fluoride
sensors:
One test container with calibration
n
solution
Incorrect calibration