Setting measured variables
8.1.11

mA General

Measured variable [mA general]
[mA general] measured variable,
With the
various preselected measured variable
can be selected and/or one measured var‐
iable can also be freely edited with its unit
of measure. The temperature measure‐
ment cannot be used for compensation
purposes, because the influence of the
temperature measurement on the meas‐
ured value is not known. In principle, the
settings are performed in the same way
as with the other measured variable. A
standardised calibrated signal is expected
by the controller from each connected
device
Features of the Two-
8.1.12
channel Version
Two channel version
If a second measuring channel is avail‐
able (dependent on the identity code,
channel 2), then this second measuring
channel can be configured according to
the descriptions of the first measuring
channel.
Two channel version with two identical
measured variables
If the measured variables of measuring
channel 1 and measuring channel 2 are
chosen identically, then the menu item
[Differential meas] appears in the
[Measurement] menu. The
[Differential meas] function is switched off
"ex works". The function
[Differential meas] can be activated and
[K1-K2] executed. The
the calculation
result of the calculation is displayed in the
main display 2 by pressing the
60
key. By pressing the
you jump back to the main display 1. The
limit value criteria for the
[Differential meas] can be set in the menu
[Limit values] .
key or
or
key again