Calibration Data For Orp Sensor (Cal) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller dialog daca
Table of Contents

Advertisement

Calibration
CAL ORP
Sensor calibration in buffer
Sensor voltage
The stability is:
acceptable
continue with <CAL>
Fig. 46: Display of the sensor stability ach‐
ieved
[acceptable / good / very good]
3.
The
range is displayed
ð The black part of the horizontal
bar indicates the range
detected.
4.
Continue with
CAL ORP
Buffer value
Offset
Accept with <CAL>
Fig. 47: Adjusting the buffer value
5.
Press
and use the four arrow
keys to adjust the mV value of the
buffer you are using
6.
Press
7.
Transfer the result of the calibration
into the controller memory by
pressing
ð The controller operates with the
calibration results.
78
0.1 mV
very good
good
A1029
165 mV
0.0 mV
A1030
Calibration data for ORP
9.2.3
sensor (CAL)
Correct sensor operation
Correct measuring and metering
is only possible if the sensor is
working perfectly
Refer to the sensor's operating
instructions
Remove the sensor from the in-
line probe housing and re-fit it for
calibration. To do this, refer to the
operating instructions for your in-
line probe housing
ORP sensor calibration
The ORP sensor cannot be cali‐
brated. Only an 'OFFSET' deviation
of magnitude ± 40 mV can be set and
thus compensated. If the ORP sensor
deviates by more than ± 40 mV from
the reference value, then it must be
checked in accordance with the
requirements of the sensor operating
instructions.