Calibration Of The Ph Sensor (Cal) By [Data Input] - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller dialog daca
Table of Contents

Advertisement

9.1.4
Calibration of the pH Sensor (CAL) by
Data input
Using the [Calibration of the pH sensor (CAL) by data input] calibration method, the
sensor's known data is entered into the controller. Calibration by data input is only as
accurate and reliable as the method with which the data was determined.
The sensor data must have been determined in real-time. The more up-to-date the
sensor data is, the more reliable is this calibration method.
Correct sensor operation
Correct measuring and metering is only possible if the sensor is working per‐
fectly
Refer to the sensor's operating instructions
Measuring and control behaviour of the controller during calibration
During calibration the actuating outputs are deactivated. Exception: a basic load or a
manual control variable has been set. This remains active. The measured value
output [standard signal output mA] is frozen, corresponding to its settings in the mA
output menu.
When calibration/testing has been completed successfully, all of the error checks
relating to the measured value are restarted. The controller saves all the determined
data for zero point and slope upon a successful calibration.
[Data Input]
Calibration
73