Note:
Two types of battery are available: the standard 5-hour and optional 10-hour.
Note:
Even if the monitoring system is turned off, the Battery Charge Indicator remains lit while
the battery recharges.
Note:
A full charge of a depleted battery takes more than four (4) hours for a 5-hour battery or
eight (8) hours for a 10-hour battery.
Plug the monitoring system into an AC outlet to charge the battery for a minimum of three
(3) minutes prior to turning on any monitoring system with a completely discharged
battery. When operating on internal battery, the monitoring system battery status icon
indicates the battery charge condition.
To charge the internal battery:
Connect the monitoring system to AC power to charge a low or depleted battery. Ref-
1.
erence
Verify the AC Power Indicator and Battery Charge Indicator light.
2.
Nellcor™ Pulse Oximetry Sensor Connection
3.5
WARNING:
Incorrect application or use of an SpO
the sensor too tightly, apply supplemental tape, or leave a sensor too long on one
place. Inspect the sensor site as directed in the Directions for Use to ensure skin
integrity, correct positioning, and adhesion of the sensor.
WARNING:
Do not use any other cables to extend the length of the Covidien-approved interface
cable. Increasing the length will degrade signal quality and may lead to inaccurate
measurements.
Operator's Manual
AC Power
Connection, p. 3-3.
sensor can cause tissue damage. Do not wrap
2
Nellcor™ Pulse Oximetry Sensor Connection