13.8 Digitizer

The digitizer feature provides additional stability to the measurement by increasing
the number of samples. This can be beneficial when measuring shorter length
fasteners with units set to in/mm hr (high resolution), in order to increase the
resolution. The reverse scenario also then applies to longer length fasteners, not
requiring the increased sample size/shots. The digitizer has three setting options of
1, 2, 4x, with 4x providing the highest number of samples/shots. It should be noted
that increasing the digitizer sample slows down the update speed. For shorter bolts,
this won't be as noticeable as using a 4x digitizer setting for longer bolts. The
procedure to change the digitizer setting is outlined below:
key once to activate the menu items tab. Press the
and
arrow keys multiple times to scroll through the sub
and
arrow keys multiple times to scroll through the
key to return to the measurement screen and begin
Setting the Digitizer
MiniMax Bolt Tension Monitor