24) Press the
Point Zero list until TEMP COEF is highlighted.
Note: This should be set to 55 (99 Metric) for the 3 & 6" bars supplied by
Dakota Ultrasonics. If the glass block is used, it should be set to 0. The reason
for this is because glass has a very low temperature coefficient.
25) Press the
26) Press the
value.
27) Press the
locations.
28) Repeat steps 26 & 27 until the TEMP COEF value is correctly displayed.
29) Press the
entering the TEMP COEF.
30) Press the
Point Zero list until POLARITY is highlighted.
Note: The default is AUTO, allowing the MiniMax the ability to select either
positive or negative polarity. However, the polarity can be set and fixed to a
specific polarity, positive/negative, as selected by the user.
and
arrow keys multiple times to scroll through the Two
key to display the Digits Edit Box.
and
arrow keys multiple times to scroll the highlighted
and
arrow keys multiple times to scroll the digit
key to return to the Two Point Zero list, or
and
arrow keys multiple times to scroll through the Two
MiniMax Bolt Tension Monitor
79
to cancel