DAKOTA ULTRASONICS MiniMax v2.0 Operation Manual page 61

Ultrasonic bolt tension monitor
Table of Contents

Advertisement

1) Press the
key multiple times to tab right and the
to tab left until the CAL menu is highlighted and displaying the submenu
items.
2) Press the
and
menu items until ZERO MODE is highlighted.
3) Press the
and
options until the AUTO option is displayed.
Note: The auto option will perform a two-point zero calibration using the first
and second multiple of the calibration standard. Therefore, if a 3" bar is used,
the second point will be at 6".
4) Press the
and
menu items until MEASURE ZERO is highlighted.
arrow keys multiple times to scroll through the sub
arrow keys multiple times to scroll the zero mode
arrow keys multiple times to scroll through the sub
55
MiniMax Bolt Tension Monitor
key multiple times