Bolt End Reflectors - DAKOTA ULTRASONICS MiniMax v2.0 Operation Manual

Ultrasonic bolt tension monitor
Table of Contents

Advertisement

Dakota Ultrasonics
Figure 3 Effect of lowered and raised grade marks

6.3 Bolt End Reflectors

Smooth, flat reflecting bolt ends that are perpendicular to the axis of the bolt are
required for accurate echo reception. Bolt ends may need to be cleaned, ground, etc.
to achieve the required surface.
Note: Misalignment exceeding 2 degrees can cause significant errors.
Avoid:
 Rough reflecting bolt ends. As shown in Figure 4, if the reflecting end of the bolt is
rough or curved, most of the reflected energy will be dispersed and a weak or
distorted echo will be received.
 Reflecting bolt ends not perpendicular to the axis of the bolt. Sonic energy will be
reflected toward the sidewall of the bolt, as shown in Figure 5.
 Nonperpendicular reflecting bolt ends due to bending of the bolt as shown in
Figure 6.
Figure 4 Rough reflective surface
Figure 5 Non Parallel reflecting Surface
42