Dakota Ultrasonics
1) Press the
Group List Items until CREATE NEW GROUP? is highlighted.
2) Press the
confirmation screen.
3) Press the
New Group setup and return to the DATA menu.
4) Press the
reference lengths.

8.4 Auto Zero/Calibration

The auto calibration feature uses a single calibration bar and two multiples (echo-
echo mode) of the calibration bar to perform an automatic Two-Point calibration. If a
3" calibration bar is used, it would measure at 3", and then again on the second
multiple at 6". This zero option will typically be the preferred option, as it's the easiest
option to use and for the most part is automatic in terms of operation. However,
depending on signal quality and calibration standard used, may not always work. If
valid signals cannot be acquired during the auto set routine, the MiniMax will report
"Can't find stable echo". Before eliminating this option altogether and using one of
the manual options listed below, try relocating the transducer to various positions on
the end of the calibration standard first. Since the calibration standards are
cylindrical, they can sometimes create substantial sidewall noise, based on the the
length, transducer size, and frequency. Moving the transducer to the edge of the
cylinder will often times eliminate the noise by shooting down the edge of the
standard and allowing the MiniMax to acquire a reasonable signal.
Now that the bolt group CAL DATA has been created in the previous section, we're
ready to get started with the automatic zero/calibration process. The first thing to
note is the procedures in this section demonstrate a Two-Point calibration, but a very
fast and efficient one, with minimal adjustments required from the operator.
Measuring Calibration Standard – For the First Time
and
arrow keys multiple times to scroll through the new
key to accept the group parameters, and activate the
key to save the New Group, or the
key to return to the measurement screen to begin measuring
54
key to cancel the