16) Repeat steps 14 & 15 until the PHY LEN 2 value is correctly displayed.
17) Press the
entering the PHY LEN 2.
18) Press the
list until VELOCITY is highlighted.
Note: This refers the velocity of the probe zero cal bars. Since there is a
velocity for each cal bar, use the velocity of the longest cal bar as it will be most
significant – 6" bar. If the glass block is being used, it will have only one
velocity specified for the block.
19) Press the
20) Press the
value.
21) Press the
locations.
22) Repeat steps 20 & 21 until the VELOCITY value is correctly displayed.
23) Press the
entering the VELOCITY.
key to return to the Two Point Zero list, or
and
arrow keys to scroll through the Two Point Zero
key to display the Digits Edit Box.
and
arrow keys multiple times to scroll the highlighted
and
arrow keys multiple times to scroll the digit
key to return to the Two Point Zero list, or