2) Press the
menu items until LOAD CAL CALC is highlighted.
3) Press the
screen.
Note: The regression or vector coefficient will be displayed. The closer this
number is to 1, the better the fit, or least amount of error.
4) Press the
return to the CAL menu.
5) Press the
Note: The new Load Factor and Offset values can be reviewed and edited in
the GEOM tabbed menu item.
6) Calibration is completed.
and
arrow keys multiple times to scroll through the sub
key to run the calculation and display the confirmation
key to accept the calculation, or the
key to return to the measurement screen.