Note: The two-point option allows the user to use two separate points for the
probe zero calibration.
Note: The one point option can optionally be selected if only one cal bar will be
used. Although the procedures below are described with the two-point option in
mind, they will also work in the case of a one-point calibration.
4) Press the
menu items until MEASURE ZERO is highlighted.
5) Press the
6) Press the
Point Zero list until PHY LEN 1 is highlighted.
Note: PHY LEN 1 is the actual physical length of the first calibration standard
noted on the certificate included with the bars. If the glass block is being used,
the 1" side should be used for PHY LEN 1.
and
arrow keys multiple times to scroll through the sub
key to display the Two Point Zero list box.
and
arrow keys multiple times to scroll through the Two