column B. It's also important to note that the L-REF stored in column A, also
contains all the MiniMax settings when the L-REF was initially measured.
These settings will automatically be copied to all the columns B-ZZ, for
each individual bolt. Therefore, no adjustments to the MiniMax should be
necessary.
10) If the elongation is +/- .0005" and "In Spec", Press the
current reading in the highlighted cell location, storing all the current probe
zero verification data. Repeat steps 6 - 10 for the other probe zero
calibration standards. If the probe zero is "Out of Spec" perform the
zero procedure in its entirety.