procedures in the next section demonstrate a Two-Point calibration. However, the
same procedures apply to the One-Point calibration option also. The user would
simply follow the same procedures and ignore references to the second point.
Measuring Calibration Standards – For the First Time
Note: It is assumed that the temperature sensor is connected to the MiniMax,
and semi automatic or automatic mode has been selected. The
temperature sensor should also be attached to the specific calibration bar
being measured. If an external pyrometer or other method will be used, the
temperature mode should be set to manual, and the temperature manually
entered into the gauge.
1) With the menu items still active from the previous section, press the
multiple times to tab right and the
CAL menu is highlighted and displaying the submenu items.
2) Press the
menu items until ZERO MODE is highlighted.
3) Press the
options until the TWO POINT option is displayed. Each time the arrow keys
are press, and confirmation screen will appear letting the user know that the
ZERO MODE is being changed. Pressing the
the mode to be changed.
and
arrow keys multiple times to scroll through the sub
and
arrow keys multiple times to scroll the zero mode
MiniMax Bolt Tension Monitor
key multiple times to tab left until the
key each time will allow
75
key