DAKOTA ULTRASONICS MiniMax v2.0 Operation Manual page 79

Ultrasonic bolt tension monitor
Table of Contents

Advertisement

Note: User has selected this section because the MiniMax, transducer, or
transducer cable has not changed and the elongation is outside the tolerance of
+/- .0005".
Note: This section assumes that a group CAL DATA has been created and
opened/active, as outlined in section 8.3.
1) Press the
key once to activate measure menu items. Press the
key multiple times to move right and the
until the LOG cell is highlighted.
2) Press the
key to display the Group View Box.
3) Press the
corresponding display view. Notice as the cursor is moved to a different
cell, the display will be updated with the display view saved with the reading.
4) Scroll the cursor to the next blank cell for the probe zero bolt. This will
always be a cell located in columns B – ZZ. As soon as the user has
scrolled to blank cell, the MiniMax will switch into elongation mode and use
all the data from the probe zero bolt initially stored.
5) Press the
multiple times to move right and the
the TEMP cell is highlighted.
arrow keys to scroll the stored readings and
key once to activate Hot Menu items. Press the
MiniMax Bolt Tension Monitor
key multiple times to move left
key multiple times to move left until
73
key