4 CALIBRATION PROCEDURE
⑥b.Actual load calibration
Apply an actual load, and register the value of the actual load (rated capacity value).
By performing actual load calibration, the setting value of equivalent input calibration becomes the
sensor output value (mV/V) of the time when it was executed. The load with which actual load
calibration was performed can be confirmed. However, it stays unchanged when an error occurs.
◇Actual load calibration setting method
(Example: ch1 actual load calibration)
Chapter
4
30
30
≪Setting value≫
-19999~19999 (except 0)
1.
Select setting mode 4.
Press
FNC
2.
Set "Actual load" of the sensor.
Press
four times. ⇒ "Actual load calibration"
is displayed.
* For performing actual load calibration on ch2,
switch with
.
Enter a setting value input state with
highest digit blinks. (Initial value: 100.00)
Check the capacity of the actually-applied load, and
input the value with
* For inputting a negative sign, press
Move to the lower digits with
Recheck that the load is properly applied on the sub display (ch2 side), and press
to validate it.
3.
Actual load calibration is executed.
If actual load calibration is executed normally, a
message as shown on the right-hand side is displayed.
If actual load calibration is not executed normally, the
alarm sounds and an error is displayed.
4.
If actual load calibration is executed normally, you will
go back to the setting item select state.
three times.
, and the
.
.
.
Ch1 Ch 2
H I
LO
H I
LO
H I
OK
LO
PEAK
HOLD
Ch1 Ch 2
H I
LO
H I
LO
Selected ch display
H I
OK
LO
PEAK
HOLD
Ch1 Ch 2
H I
LO
H I
LO
LO
PEAK
HOLD
H I
OK
Ch1 Ch 2
H I
LO
H I
LO
H I
OK
LO
PEAK
HOLD
Ch1 Ch 2
H I
LO
H I
LO

Advertisement

Table of Contents
loading

Table of Contents