■Connection

Example)F350
 XM3D-0921 or equivalent
1
2
3
4
5
6
7
8
9
* This connection diagram shows cabling applicable to the case where your PC is DTE (data
terminal equipment). For connection with DCE (data circuit-terminating equipment), such as a
modem, use straight type cabling.
* Prepare cabling after reconfirmation of the connector shape and signal lines (pin assignments) of
the equipment you use.

■Communication mode

《Setting value》
0:Send/receive
1:Continuous send
2:Print send mode
◇Communication mode setting method
1.
2.
3.
71
RXD
TXD
DTR
GND
RTS
CTS
This mode performs communication by a command from the host computer.
In this mode, you can read out the indicated value, status, set values and write in
set values.
This mode continuously transmits the indicated values and the status.
When the indicated value is printed (a print command is output to the SI/F), the
indicated value and status are sent from the F350 to host.
Select setting mode 5.
Press
FNC
Set "Communication mode."
Press
once. ⇒ "Communication mode" is
displayed.
Press
, and present set value blinks.
Input the setting value with
Press
to validate it.
Press
twice to go back to the indicated value display.
ESC
four times.
. (Initial value : 0)
9 OPTION INTERFACE
PC etc.
D-sub9pin
1
CD
2
RXD
3
TXD
4
DTR
5
GND
6
DSR
7
RTS
8
CTS
9
RI
Ch1
Ch 2
H I
H I
LO
LO
PEAK
H I
OK
Ch1 Ch 2
H I
H I
LO
Chapter
9
LO
HOLD
LO