Sub Numerical Display Section - Unipulse F350 Operation Manual

Digital indicator
Table of Contents

Advertisement

11 SUPPLEMENTS

■Sub numerical display section

-LoAd
LoAd
ALA
-ALA
oFL1
oFL2
cErr2
cErr3
cErr4
cErr5
cErr6
dZ Err
Chapter
CALZE
11
CALSP
CALEq
98
98
Beyond the signal input
range -3.2mV/V
Beyond the signal input
range +3.2mV/V
Exceeds the alarm HI limit
Below the alarm LO limit
Below 19999
Exceeds -19999
Beyond the zero calibration
range in the positive
direction
Beyond the zero calibration
range in the negative
direction
Beyond the rated output
value in the positive
direction/negative direction.
Beyond the rated capacity
value.
Rated capacity value = 0
Rated output value
0.000mV/V at the time of
equivalent input calibration
Rated output value in the
range of -0.005 to
+0.005mV/V at the time of
actual load calibration
Digital Zero limit error
Zero calibration in progress
Actual load calibration in
progress
Equivalent input calibration
in progress
The input electric signal greatly exceeds the signal
input range. This also occurs when an excessive load
is applied to the sensor or the input terminal is opened
due to a break in the cabling, etc.
This occurs when a larger load than expected is
applied to the sensor or calibration is performed below
the measurement range. Check the calibration and
measuring object.
This occurs when a larger load than expected is
applied to the sensor, or calibration is performed below
the measurement range. Check the calibration and
measuring object.
The signal input range is greatly exceeded when zero
calibration is performed. Check for an excessive load
applied to the sensor and for breaks.
The signal input range is greatly exceeded when
calibration is performed. In the case of equivalent input
calibration, check the rated output value, and in the
case of actual load calibration, check for an excessive
load applied to the sensor and for breaks.
Also, Check the setting of rated capacity value (display
value) sets it exceeding "±19999".
Check the setting of rated capacity value (display
value) for being "0" when calibration is performed.
Check the rated output value for being 0.000mV/V
when equivalent input calibration is performed, and the
load for being too light in weight when actual load
calibration is performed.
When the range of zero point correction (deviation from
the zero calibration registered point) by Digital Zero or
Zero Tracking exceeds the setting value of Digital Zero
limit, Digital Zero limit error occurs. Perform Digital
Zero, etc., in a proper range.
These are not errors. They indicate that each
calibration is in execution.