Unipulse F350 Operation Manual page 38

Digital indicator
Table of Contents

Advertisement

①Releasing the calibration protect
Turn off the calibration protect that inhibits calibration.
 ≪Setting value≫
◇Calibration protect setting method
1.
2.
3.
②Excitation voltage setting
Select the bridge voltage (excitation voltage) supplied to the strain gauge sensor.
The bridge voltage can be selected from 2.5V and 10V.
After this setting, make sure to perform calibration.
≪Setting value≫
* Common to ch1 and ch2.
③Decimal place setting
Set the decimal place for load-value-related display, setting items, etc.
≪Setting value≫
* Calculated values are also synchronized with this setting.
25
0:OFF
Enables rewriting of setting values.
1:ON
Disables rewriting of setting values.
Select setting mode 9.
Press
five times.
FNC
Set "Calibration protect."
Press
once. ⇒ "Setting protect" is displayed.
Press
calibration protect blink.
Input the setting value with
(Initial value: 0)
Press
to validate it.
Press
twice to go back to the indicated value
ESC
display.
0:2.5V
1:10V
0:None
1:0.0
2:0.00
3:0.000
4 CALIBRATION PROCEDURE
twice to make the digit of the
.
Ch1 Ch 2
H I
LO
H I
LO
H I
OK
LO
PEAK
HOLD
Ch1 Ch 2
H I
LO
H I
LO
《Setting value》
0 0 0
Calibration protect
 0 : OFF 1 : ON
System protect
 0 : OFF 1 : ON
Work protect
 0 : OFF 1 : ON
Chapter
4
25