4 CALIBRATION PROCEDURE
④Min. scale division setting
Set the minimum unit (scale interval, scale division) of measurements.
Chapter
4
* Calculated values are also synchronized with this setting
◇Excitation voltage, decimal place, and min. scale division setting method
26
26
The maximum excitation voltage value of the strain gauge sensor connected to the
F350 should always be larger than the set bridge voltage.
If the excitation voltage of the F350 is larger than the maximum excitation voltage
value of the strain gauge sensor, the strain gauge sensor may heat or break.
≪Setting value≫
0:1
1:2
2:5
3:10
4:20
5:50
6:100
1.
Select setting mode 4.
Press
FNC
2.
Set "Excitation voltage," "Decimal place," and "Min.
scale division."
Press
once. ⇒ "Calibration function select"
is displayed.
Press
, and the highest digit blinks.
① Select excitation voltage.
Input the excitation voltage with
move to the lower digit with
② Select decimal place.
Input the decimal place with
move to the lower digit with
③ Select min. scale division.
Input the min. scale division with
validate it with
3.
Press
twice to go back to the indicted value display.
ESC
CAUTION
three times.
, and
. (Initial value: 0)
, and
. (Initial value: 2)
, and
.(Initial value: 0).
Ch1 Ch 2
H I
LO
H I
H I
OK
LO
PEAK
HOLD
Ch1 Ch 2
H I
LO
H I
《Setting value》
0
2
0
Min. scale division
0 : 1
4 : 20
1 : 2
5 : 50
2 : 5
6 : 100
3 : 10
Decimal place
0 : None
1 : 0.0
2 : 0.00
3 : 0.000
Excitation voltage
 0 : 2.5V
 1 : 10V