9 OPTION INTERFACE
9-3.

D/A converter output

A D/A converter is provided for obtaining analog output synchronized with the indicated value of
the F350.The analog output ranges are -10 - +10V output and 4 - 20mA constant-current output.
By using the D/A zero setting and D/A full scale setting functions, analog output can be obtained
between zero (0V, 4mA) and full scale (10V, 20mA) with respect to the predetermined digital value.
The output circuit and the main unit circuit are isolated.
The resolution is 1/10000 with respect to the ranges of -10 to +10V and of 4 to 20mA, and the
conversion speed follows the A/D conversion speed.
The maximum voltage output range are about ±11V, the maximum current output range are about
2.4 - 21.6 mA

■Connection

For the output terminals of the D/A converter option, a cage clamp type terminal block is used.
Chapter
Make connections by the following procedure.
9
78
78
DAV (D/A converter voltage output)
VOUT
GND
DAI (D/A converter current output)
IOUT
GND