2 INSTALLATION & CONNECTION

■Analog input/output connector connection

Connect a strain gauge sensor or voltage output.
Make connections according to the silk-screen printing on the rear panel of the F350.
●Pin Assignment
Chapter
2

■How to remove the terminal block

10
10
12
SHIELD
11
-MONITOR(ch2)
10
+MONITOR(ch2)
9
-MONITOR(ch1)
8
+MONITOR(ch1)
7
-SIG(ch2)
6
+SIG(ch2)
5
-SIG(ch1)
4
+SIG(ch1)
3
-EXC*
2
+EXC*
1
SHIELD
Compatible connector: ETB42-12P(OSADA CO.,Ltd.) or equivalent
(Connector optional type:CN81)
Before connecting a strain gauge sensor, make sure to check the setting of excitation
voltage.
For the setting of excitation voltage, see "4-2 ②Excitation voltage setting" on page
25.
1.
Loosen the screws (two) with a screwdriver.
2.
Remove the terminal block by giving it a
strong pull.
Attention
When installing the terminal block to the F350
body, check its orientation. (See the illustration at
the right.)
  * EXC is common to ch1 and ch2.
CAUTION
Loosen.
(Turn
counterclockwise
.)
Insert side
Right side