■Mode 1; work setting (comparison/hold)
Setting item
No.
Name
Comparison function select
1
Near Zero
2
Hysteresis
3
Calculation hold function select
4
Detection time
* This setting is required only when
5
the hold section is "external signal +
time."
Average sample number
6
* This setting is required only when
the hold mode is "average."
7
8
9
89
H I
OK
LO
PEAK
Initial
Setting range
value
0 0
0
0
Undefined
Undefined
Calculation comparison OK
output condition
0: Calculated value OK only
1: Calculated value OK &
ch1 value OK
2: Calculated value OK &
ch2 value OK
0000
3: Calculated value OK &
ch1 value OK &
ch2 value OK
Calculation comparison timing
0: All time
1: Stable
2: Near Zero OFF
3: Stable and Near Zero OFF
4: Hold
100
0~99999
0
0~9999
0 0
0
1
Auto reset
0 : OFF
1 : ON
* This setting is required when the
hold section is "external signal"
or "external signal + time."
Hold section
0 : All section
1 : External signal
2 : External signal + time
* This setting is required only when the
hold mode is other than "0: OFF."
* When the hold mold is "1: Sample,
0001
" settings are available up to
"1: External signal."
Hold mode
 0 : OFF
 1 : Sample
2 : Peak
3 : Bottom
4 : Average
Calculation mode
 0 : Ch1+Ch2
 1 : Ch1-Ch2
 2 : |CH1-CH2|
 3 : High select
 4 : Low select
0.01 to 9.99 [sec]
1.00
1 to 999 [times]
1
11 SUPPLEMENTS
HOLD
Opera
Setting
by ch
tion
Protec
NOV
Refer
t
RAM
ence
P48
P49
P47
P45
P44
P50
Chapter
P57
11
P53
89