Cleaning the
Archive Processor
(periodic cleaning)
Cleaning the
developer racks and
tank
A-62051 May 2008
Clean the developer racks with Kodak Developer System Cleaner and
Neutralizer, or equivalent, every time the chemical tanks are changed
using automatic replenishment (every third time with manual
replenishment) to remove silver and chemical deposits.
NOTE: Discharge, treatment, or disposal of spent solutions may
be subject to local, state, or federal laws. Contact
appropriate authorities to determine the requirements that
apply to the use of this product.
CAUTION: Follow the safe-handling instructions printed on the
bottle label.
1. Mix the Kodak System Cleaner.
• Start with 750 ml of water and add 125 ml of cleaner part "A"
concentrate and add 125 ml of cleaner part "B" concentrate then
mix until the solution is uniform.
or
• 6 parts of water add 1 part of cleaner "A" and 1 part of cleaner "B"
then mix until the solution is uniform.
NOTES:
• Do not mix developer system cleaner directly with neutralizer.
• Improper mixing and handling, such as allowing the cleaner to come
in contact with residual developer or the other processing solutions,
can form irritating sulfur dioxide fumes. Use adequate ventilation.
2. Mix neutralizer
• Start with 875 ml of water, add 125 ml of neutralizer concentrate
then mix until the solution is uniform
or
• 7 parts of water, add 1 part of neutralizer concentrate then mix
until uniform
3. Disconnect the power from the wall outlet.
4. Lift the processing section cover.