Making The Plumbing Connections - Kodak X-Omat 270 RA Installation Instructions Manual

Hide thumbs Also See for X-Omat 270 RA:
Table of Contents

Advertisement

Making the Plumbing Connections

Connecting the Water Supply
[1]
Connect the incoming water supply to the WATER SUPPLY INLET. Figures 20 and 21 show the
complete water-to-PROCESSOR layout and connecting fittings.
• A mixing valve is not required for the PROCESSOR if the incoming water temperature is
between 4° and 29°C (40° and 85°F). However, tempered water is recommended for mixing
chemicals and cleaning the processing TANKS.
• The PROCESSOR has an internal 12.7 cm (5 in.) water gap in the WASH RACK. A Check Valve (or
Vacuum Breaker) should not be necessary; however, check and observe local codes.
• The PROCESSOR accepts water pressure from 170 - 690 kPa (25 - 100 psi). Higher pressures may
cause the WATER INPUT SOLENOID to malfunction. A water pressure of 410 kPa (60 psi), or higher,
may increase noise. If necessary, install a Regulator.
• A 50 micron Water Filter is required (not supplied by Kodak) in the input line.
Check Valve
Service Hose
(long enough
for cleaning
processor)
Dashed lines represent
darkroom side.
Service line must
pass through wall
2B6843
NOTE
Water Filter
Thermometer Part No. 458928
Locate Union
close to Processor
Figure 20 Connecting the Water Supply
Shutoff Valve
Hot and Cold
Water Supply
Processor
Receive End
Stand
H104_0083DCA
H104_0083DA
25