Key Operator Instructions

Calibrating the Replenishment System

5–70
Important
Calibrate the Replenishment System every 3 months, or whenever you add new
chemicals to the Processor.
Calibrate the Replenishment System for all process cycles in 1 operation that
determines the actual rate of solution flow through the Replenisher Pumps. The
operator will measure the volume of solution pumped during a set period of time, then
use the Display Panel to enter the measurement into the Microprocessor. The
Microprocessor computes the rate of solution flow through the Pump, then adjusts the
period of time that the Pump must operate to match the Replenishment volume set by
the operator.
Note
The volume actually measured during this procedure is not the volume delivered for a
35 x 43 cm sheet of film.
1. Press GO TO SETUP.
PROCESSOR READY
127.4 F = DRYER TEMP
DISPLAY
TIME/
FIX TEMP
DATE
H186_8522AC
2. Enter the 4-digit access code.
1
H186_8523AC
95.0 F = DEV TEMP
GO TO
DONE/
SETUP
RETURN
2
3
4
CANCEL
REQUEST
STD
16FEB01 – 7E6174