Preselecting basic settings

5.2
5.2.1

Changing the basic settings and start settings

5.2.1.1

Reprogramming the kV/mA values

The factory-programmed kV/mA combinations for each preselected
exposure program and for each individual patient symbol key can be
overwritten (reprogrammed).
IMPORTANT
If no entry is made for a period of time exceeding 5 seconds during
the programming procedure, the programming mode is automatically
terminated without saving any of the changes previously entered.
1. Select the exposure program for which the values are to be
B
changed. Press the program selection keys (A) forward/backward.
Ä The desired program is displayed.
2. Briefly press the Memory key (C).
Ä The LED above the Memory key lights up.
3. Select the patient symbol corresponding to the kV/mA value you
want to change. Press the desired patient symbol key (D).
Ä The LED above the patient symbol key lights up.
4. Set the new kV/mA value. Push the kV/mA keys (B) forward and
backward.
Ä The desired kV/mA value is displayed.
5. Save the setting. Briefly press the Memory key (C).
C
Ä The LED above the Memory key briefly flashes and then
switches off. The program display jumps back to P1 for PAN
programs or C3 for Ceph programs.
Ä The new kV/mA value is programmed.
5 Operation
5.2 Preselecting basic settings
89