Dentsply Sirona Orthophos E Ceph Operating Instructions Manual page 57

Hide thumbs Also See for Orthophos E Ceph:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Orthophos E
G
F
E
D
A
B
H
67 11 001 D3352
D3352.201.12.01.02    03.2019
7. Align the patient with the central laser line G.
Ä The laser line reflects in the center of the patient's anterior teeth
or the middle of the face (mid-sagittal).
8. Align the patient's head according to the Frankfurt horizontal plane
E
(E).
9. Adjust the height of the light localizer using the slider (F).
Ä The laser line reflects on the upper edge of the outer ear canal.
10. Correct the patient's head inclination as necessary. Briefly press the
up A and down B height adjustment keys.
Ä The laser line reflects on the lowest point of the lower eye
socket edge.
11. Press the forehead support adjustment key (H).
Ä The forehead supports stop moving automatically when they
come into contact with the patient's forehead.
Ensure that the patient's head does not move backward when
the forehead support is put in place.
12. Check the patient's position and make any final corrections as
necessary.
13. Swivel the control mirror back into place. by pressing the right
recess on the touchbar (D).
Ä The patient can see himself in the control mirror.
14. Instruct the patient to exhale, place his tongue against the roof of
his mouth, and hold this position until the end of the exposure.
Ä The patient is positioned in the unit.
IMPORTANT
Note the information on the two-part exposure programs in the
"Releasing the exposure" chapter, see "Two-part exposure
programs".
5 Operation
5.1 Acquiring the X-ray image
57