Yy - Locality - Dentsply Sirona Orthophos E Ceph Operating Instructions Manual

Hide thumbs Also See for Orthophos E Ceph:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Orthophos E
67 11 001 D3352
D3352.201.12.01.02    03.2019
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1. Acknowledge the error message.
Ä The error no longer occurs.
2. If the error remains, switch the unit off and on again. NOTE!
 Observe waiting period!
Ä The error no longer occurs.
3. If the error persists, contact your Customer Service.
E7 – Serious system error
The error occurs spontaneously and without a corresponding operation.
1. Switch off the unit.
2. Contact your Customer Service immediately.
Ä The unit is functional.
7.3.2
yy – Locality
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
06 – Tube assembly
71 – Multipad user interface
10 – Central control DX 11; system hardware
11 – Central control DX 11; system software
12 – Central control DX 11; central CAN bus fault
13 – Central control DX 11; DX11, DX1 periphery (motor system of
stand, sensor system of stand)
14 – Central control DX 11; digital extension (HSI, network, etc.)
15 – Central control DX 11; configuration (wrong software, wrong
module constellation, etc.)
16 – Central control DX 11; Zero Management
20 – Central control DX 11; Framegrabber application
22 – Central control DX 11; 2D Imaging System
23 – Central control DX 11; 3D Imaging System
42 – Remote
61 – Diaphragm control
81 – Sensor
91 – Ceph digital
7 Malfunctions
7.3 Error description
99