8 Program values

8.3 Dose information

Program
Maximum
effective radia-
tion time
P1
14.2s
P1L
8.0s
P1R
8.0s
P1A
14.2s
P1C
14.2s
P10
11.6 s
P12
4.9s
BW1
8.8s
TM1.1+TM1.2
6.4+6.4s
S1
14.4s
MS1
21.7s
104
Dose information
8.3
The dose area product for the parameter combinations proposed by
Dentsply Sirona has been calculated already. The DAP value can be
used without any further calculations.
8.3.1

Dose area product parameters for Panorama images

The dose area product (DAP) values were measured in a CT ionization
chamber.
Dose level series index 4A (8 mA/16 mA series)
Specification of the dose area product (DAP/absorbed dose) for
panoramic views, temporomandibular joint (TMJ) views and sinus views
Factory-programmed values
DAP
2
mGycm
kV/mA
kV/mA
62/8
64
64/8
62/8
36
64/8
62/8
36
64/8
62/8
64
64/8
62/8
64
64/8
62/8
33
64/8
71/8
29
77/7
62/8
20
64/8
68/8
69
71/8
71/8
85
77/7
73/8
135
77/7
Operating Instructions Orthophos E
DAP
DAP
2
mGycm
mGycm
kV/mA
68
69/15
147
39
69/15
83
39
69/15
83
68
69/15
147
68
69/15
147
35
69/15
75
30
80/14
63
21
69/15
46
75
73/15
147
87
80/14
185
131
80/14
280
D3352.201.12.01.02    03.2019
Dentsply Sirona
DAP
2
2
mGycm
kV/mA
73/15
164
73/15
92
73/15
92
73/15
164
7315
164
73/15
83
84/13
63
73/15
51
77/14
151
90/12
192
84/13
282
67 11 001 D3352