Operating Instructions Orthophos E
Program
Maximum
effective radia-
tion time
P1
14.2s
P1L
8.0s
P1R
8.0s
P1A
14.2s
P1C
14.2s
P10
11.6 s
P12
4.9s
BW1
8.8s
TM1.1+TM1.2
6.4+6.4s
S1
14.4s
MS1
21.7s
67 11 001 D3352
D3352.201.12.01.02    03.2019
Dose level series index 2A (8 mA series)
Specification of the dose area product (DAP/absorbed dose) for
panoramic views, temporomandibular joint (TMJ) views and sinus views
Factory-programmed values
DAP
2
mGycm
kV/mA
kV/mA
62/8
64
64/8
62/8
36
64/8
62/8
36
64/8
62/8
64
64/8
62/8
64
64/8
62/8
33
64/8
71/8
29
77/7
62/8
20
64/8
68/8
69
71/8
71/8
85
77/7
73/8
135
77/7
DAP
DAP
2
mGycm
mGycm
kV/mA
68
68/8
77
39
68/8
43
39
68/8
43
68
68/8
77
68
68/8
77
35
68/8
39
30
80/7
32
21
68/8
24
75
73/8
79
87
80/7
94
131
80/7
142
8 Program values
8.3 Dose information
DAP
2
2
mGycm
kV/mA
73/8
89
73/8
50
73/8
50
73/8
89
73/8
89
73/8
45
84/6
30
73/8
28
77/7
77
90/6
98
84/6
134
105