Dentsply Sirona Cavitron 300 Series Directions For Use Manual

Dentsply Sirona Cavitron 300 Series Directions For Use Manual

Ultrasonic scaling system
Hide thumbs Also See for Cavitron 300 Series:

Advertisement

Cavitron
®
Ultrasonic Scaling System
Directions For Use
Please read carefully and completely before operating unit.
300 Series

Advertisement

Table of Contents
loading

Summary of Contents for Dentsply Sirona Cavitron 300 Series

  • Page 1 Cavitron 300 Series ® Ultrasonic Scaling System Directions For Use Please read carefully and completely before operating unit.
  • Page 2: Table Of Contents

    Table of Contents Product Overview Technical Support Supplies & Replacement Parts 1. Indications For Use 2. Contraindications 3. Warnings 4. Precautions 5. Adverse Reactions 6. Infection Control 7. Installation Instructions 8. Cavitron 300 Series Ultrasonic Scaler Description ® 9. System Setup, Operation and Techniques for Use 10.
  • Page 3: Product Overview

    For dental use only. some types of electronic equipment might interfere with the Product Overview operation of the device. Although no instance of interference has ever been reported to Dentsply Sirona, we recommend that The Cavitron 300 Series Ultrasonic Scaling System offers ®...
  • Page 4: Adverse Reactions

    6. Infection Control verify that all components and accessories are included: 6.1 General Infection Control Cavitron 300 Series Ultrasonic Scaler with Detachable Featherweight Handpiece Cable • For operator and patient safety, carefully practice the infection 2. Water Line Assembly (Blue) with Filter and Quick Disconnect control procedures detailed in the Cavitron Systems Infection Control Procedures booklet accompanying your system.
  • Page 5 7.8 Charging the Foot Pedal Ensure the foot pedal power is ON. With the unit ON, plug the Auxiliary foot pedal cable into the foot pedal and the USB port on the front of the unit. Allow to charge for up to 4 hours for a full Main Power Switch battery charge.
  • Page 6: Cavitron ® 300 Series Ultrasonic Scaler Description

    8. Cavitron 300 Series Ultrasonic Scaler ® 8.1 System Controls Diagnostic Display See section 8.2 for more details. Purge Purge mode is used to flush water lines. Runs automatically Main Power ON/OFF Switch for 2 minutes (Page 8). ON/OFF Switch located located on the rear of the Rinse system.
  • Page 7 8.2 Diagnostic Display Indicators and Control Service Main Screen Cable or handpiece issue. Unit overheating. Stop use and contact Cavitron Care. Filter Icon will appear when water filter needs to be replaced. Once replaced, press and hold to reset indicator. Purge Purge mode is used to flush water lines.
  • Page 8 8.3 Handpiece/Cable 360 Rotating Handpiece Nose Lavage Control Featherweight, Detachable Handpiece Cord To rotate the insert, place fingers Turn the Lavage Control to select flow rate Lightweight handpiece cable easily on the nose of the handpiece and during system operation. Flow rate is based detaches from the unit, also detaches from rotate to desired position.
  • Page 9: System Setup, Operation And Techniques For Use

    3. Lavage (Water) Filter, 10/Pack, Part Number 90158 down to activate the system.) Check water spray to verify fluid For detailed information, contact your local Dentsply Sirona is reaching the working end of the insert tip. Adjust the water Representative or authorized Dentsply Sirona Distributor.
  • Page 10: System Care

    10. System Care 10.2 Weekly Maintenance It is strongly recommended that this system be disinfected It is recommended that you perform the following maintenance by chemically flushing the waterlines with a 1:10 Sodium procedures. Hypochlorite solution (NaOCl) at the end of each week. This 10.1 Daily Maintenance can be accomplished by connecting this device to the Cavitron Start-Up Procedures at the beginning of the day:...
  • Page 11: Troubleshooting

    For technical support and repair assistance call Dentsply Sirona Cavitron Care Factory Certified Service at 1-800-989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For areas outside the U.S., contact your local Dentsply Sirona representative. English - 11...
  • Page 12: Warranty Period

    WARRANTY STATEMENT Dentsply Sirona Cavitron products are designed expressly for dental use, and this warranty is not applicable to other uses. The products are warranted against defects arising from faulty materials or workmanship from date of purchase for the period stated above. Parts will be repaired or replaced at the discretion of Dentsply Sirona.
  • Page 13: Specifications

    13. Specifications Electrical Voltage Continuous (100-240 VAC) Current 1.0 Amperes, Maximum Phase Single Frequency 50/60 Hertz Water Pressure 20 to 40 psig (138 to 275 kPa) Water Temperature < 25°C (77°F) Water Flow Rate Minimum Setting (CCW) < 15 ml/min Maximum Setting (CW) >...
  • Page 14: Symbol Identification

    14. Symbol Identification AC POWER AC Power TYPE B APPLIED PART EQUIPMENT TYPE B APPLIED PART EQUIPMENT PROTECTIVE EARTH (GROUND) PROTECTIVE EARTH (GROUND) Footswitch not for operating theatres IPX1 Protection Class- IPX1 IPX1 Classification of ingress of water Consult Instructions for Use Consult Instructions for Use AC Power Switch (0 = Off, | = On)
  • Page 15: Classifications

    16. Disposal of Unit U.S. - Dispose of the system components in accordance with state and local laws. 17. Electro-Magnetic Compatibility Precautions Document can be viewed on the Dentsply Sirona website https://www.dentsplysirona.com under the Cavitron 300 Series Ultrasonic ® Scaling System Resources.